Steam jet test chamber

Technical parameters:
All parts of the steam jet test chamber in contact with the medium are rust-proof and can be operated with deionized water.
Pump body:
Pump - with direct drive motor and ceramic plunger, high safety, low wear, safety valve, including control motor protection injection time can be set by a timer in the display, lasting from 1 second to 9 minutes
The pressure can be adjusted externally via a pressure control valve. Pressure depends on the spray - see manufacturer's specifications
Technical data: Flow rate :14.8 L/min
Working pressure P:30-140bar can be adjusted by external pressure regulator
Motor N:6.5 kW, 400V Plug :32 amps, 5 poles
Water supply :2-3 bar, 15°C
Nozzle distance :0-450mm
Test Angle :90° -45 ° -30 °+/ -2 °, door with inspection glass, exhaust bore 102mm, fitting, discharge port with dirt filter.
Bottom frame: stainless steel base with housing and 4 guides.
Sterilization water change program: Execute the water change program every 7 days, automatic cycle sterilization water change program - through manual water change, the water in the tank is automatically heated to 68 ° C, when the tank is heated to 68 ° C temperature water is sprayed out through the nozzle. Once the tank reaches 70 ° C, the automatic cycle of the sterilization and water change procedure will be stopped, and after completion, the tank must be cooled to the operating temperature.
Door lock device: The door is automatically electromagnetic locked during pump operation. The door cannot be opened while the pump is running
Heating container: heating power 9,0 kW, maximum temperature 80°C +/-2, digital display, preheating time :1 hour