Composting Biodegradation Tester

The biodegradation rate is calculated as the percentage of the amount of carbon dioxide released during the biodegradation of the sample relative to its theoretical carbon dioxide release amount.
Working Principle
This experimental system simulates the biodegradation process of bio-based materials under aerobic composting conditions. In an aerobic environment, microorganisms use bio-based materials as a carbon source, consume oxygen, and produce inorganic substances such as carbon dioxide and water. The carbon dioxide generated is absorbed by an alkaline absorbent, and the resulting increase in weight represents the amount of carbon dioxide released during the biodegradation of the sample.
Applications
1. Meets the testing requirements of the GB/T 19277.2 standard.
2. Suitable for biodegradability testing using the composting method.
3. Measures carbon dioxide generation using the gravimetric method.
4. Suitable for determining the biodegradability of bio-based materials, plastics, and other products.
Technical Specifications
1. Working Conditions
No. | Parameter | Specification |
1.1 | Power Supply Voltage | AC 220 V ±10%, 50 Hz, single-phase |
1.2 | Operating Temperature | 15–30°C |
1.3 | Relative Humidity | ≤80% |
1.4 | Continuous Operation | Capable of continuous operation for extended periods |
2. Oxygen Supply System
No. | Parameter | Specification |
2.1 | Pre-treatment Decarbonization System | Available |
2.2 | Gas Supply Method | Continuous aeration for oxygen supply, with a built-in gas supply pump |
2.3 | Gas Pump Control | 9 independent channels, each with separate control |
2.4 | Flow Control Method | PLC control with automatic flow adjustment according to the set value |
2.5 | Flow Measurement | 9 electronic flow meters for flow measurement |
2.6 | Gas Flow Adjustment Range | 0–300 mL/min |
2.7 | Flow Control Accuracy | ±1% |
2.8 | Flow Control and Measurement | Operate independently |
3. Temperature Control System
No. | Parameter | Specification |
3.1 | Temperature Control Method | Air bath |
3.2 | Set Temperature Range | Room temperature to 90°C, in increments of 0.1°C |
3.3 | Temperature Control Accuracy | ≤±0.1°C |
3.4 | Chamber Door Design | Top-mounted sliding door for convenient operation |
4. Reactor System
No. | Parameter | Specification |
4.1 | Number of Channels | ≥9 |
4.2 | Reactor Volume | ≥2,000 mL |
4.3 | Reactor Design | Double-layer design |
4.4 | Water Addition Method | Manual |
4.5 | Stirring Method | Manual |
4.6 | Reactor Bottle Material | High-borosilicate hard glass |
4.7 | Gas Supply Method | Bottom gas supply |
5. Condensation and Dehumidification System
No. | Parameter | Specification |
5.1 | System Components | Low-temperature thermostatic bath, fully automatic condensation bottles, and dehumidification bottles |
5.2 | Channel Configuration | Each channel is equipped with its own condensation bottle and dehumidification bottle |
5.3 | Condensation Method | Water-bath condensation |
5.4 | Condensation Temperature | As low as 5°C |
5.5 | Drainage Method | Automatic discharge of condensate at preset intervals |
5.6 | Dehumidification Bottle Design | Double-layer design |
5.7 | Connection Method | Condensation bottles and dehumidification bottles connected in series |
6. Absorption and Weighing System
No. | Parameter | Specification |
6.1 | Number of Channels | 9 |
6.2 | Gas Inlet Method for Absorption Bottles | Bottom gas supply |
6.3 | Number of Absorption Bottles | 2 per channel, 18 in total |
6.4 | Absorption Bottle Connection Method | Series connection |
6.5 | Absorption Bottle Design | Double-layer design |
6.6 | Absorption Bottle Capacity | 500 mL |
6.7 | Weighing Method | Manual weighing |
6.8 | Weighing Frequency | Once per day |
7. Software System
No. | Parameter | Specification |
7.1 | Software Control System | Independently developed software system with proprietary intellectual property rights |
7.2 | Data Acquisition and Display | Real-time data acquisition and generation of real-time data curves |
7.3 | Supported Operating Systems | Professional editions of Windows 7, Windows 8, Windows 10, and other compatible operating systems |
7.4 | Data Storage Method | Real-time data storage during experiments |
7.5 | Cloud Platform | Supports remote data access via a cloud platform |
Product Features
1. Determines the amount of carbon dioxide released by the sample using the gravimetric method.
2. Features a 12-channel testing system for high experimental efficiency.
3. Uses air-bath temperature control, with a temperature adjustment accuracy of up to ±0.1°C.
4. Features an upward-opening sliding door design for the temperature-controlled chamber, facilitating operation.
5. The software collects data in real time. Data can be stored and exported, and cloud platform monitoring is supported.
Leave Message Get Price