Masks Bacterial Filtration Efficiency (BFE)Tester
Mask Bacterial Filtration Efficiency Tester is applied to test the filtration efficiency of mask, can be used in metering calibration department, scientific research institution, mask manufacturer and other related departments.
Standards:
YY 0469-2004 Technical requirements for surgical mask
BS EN 14683-2014 Medical face masks - Requirements and test methods
ASTM F2100-2019 Standard Specification for Performance of Materials Used in Medical Face Masks
ASTM F2101 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
Feature:
1.Bacterial Filtration Efficiency Tester Of Mask: Negative pressure testing system, which can protect operator’s safety.
2.Peristaltic pump was built in the negative pressure cabinet, A & B two routes 6 level Andersen sampling head.
3.The flow speed of peristaltic pump can be set.
4.The spray flow of bacterial liquid inside the special microbial aerosol generator can be set.
5.BFE tester good atomization effect.
6.Embedded high speed industrial microcomputer control.
7.10.4 "industrial high brightness color touch screen display.
8.USD connector, support USB data archived
9.Mask BFE Tester high brightness light installed in the cabinet.
10.Switching type glass door in front of the machine, easy to observe by the operator.
Specification:
Key Specification | Specification Range | Resolution | Max. permissible errors |
A Route Sampling Flow | 28.3L/min | 0.1L/min | Within ± 2.0% |
B Route Sampling Flow | 28.3L/min | 0.1L/min | Within ±2.0% |
Spray Flow | (8~10)L/min | 0.1L/min | Within ±5.0% |
Peristaltic Pump Flow | (0.006~3.0)mL/min | 0.001ml/min | Within ±2.0% |
Front Pressure of A Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.0% |
Front Pressure of B Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.0% |
Front Pressure of Spray Flowmeter | (0~300)kPa | 0.1kPa | Within ±2.0% |
Negative Pressure of the Chamber | (-90~-120)Pa | 0.1Pa | Within ±1.0% |
Working Temperature | 0-50℃ | ||
Negative Pressure of the Chamber | >120Pa | ||
Data Save Ability | >500000 sets(Scalable capacity) | ||
High Efficiency Particulate Air Filter Properties | ≥99.995%@0.3μm, ≥99.995%@0.12μm | ||
Median diameter of aerosol generator mass | Average diameter: (3.0±0.3)μm; Geometric Standard Deviation≤1.5 | ||
Double Routes 6-lever Andersen Sampling apparatus |
LevelⅠ>7μm; Level Ⅱ: (4.7~7)μm; Level Ⅲ: (3.3~4.7)μm; Level Ⅳ: (2.1~3.3)μm; Level Ⅴ: (1.1~2.1)μm; Level Ⅵ: (0.6~1.1)μm |
||
Size of Aerosol Chamber | 600×80×3mm (Length×Diameter×Thickness) | ||
Total number of positive quality control sampler particles | (2200±500)cfu | ||
Negative Chamber Flow Speed | ≥5m3/min | ||
Size of Main Machine |
1470×790×2100mm (L×W×H) |
||
Power Supply | AC220V±10%, 50Hz | ||
Noise of the Tester | <65dB(A) | ||
Weight | ≈150kg | ||
Power Consumption | <1500W |
Note:QINSUN always been committed to innovation and improve product performance and functionality, based on the reasons, the product specifications will be changed accordingly. The above without prior notice, the Company reserves the right to amend the final.